The 99 Ultralites provide a wide area of protection against x-ray. Their high curvature provide adequate lateral protection which eliminates the need for side shields. The 99 comes in 2 sizes and an optional soft rubber nose bridge for additional support and comfort (99R). (Blue only available in regular size and Fuchsia only available in small).
FEATURES
- Rubberized temples for comfort and grip
- High curvature for maximum field of view and x-ray protection
- Optional rubberized nose bridge
- Available in 2 sizes
99 ULTRALITE CHARACTERISTICS
- Frame Type: Wraparound
- Front Lens: 0.75mm Pb Glass
- Weight: 63g (regular) | 59.5g (small)
- Frame Size: 61-18-140 (regular) | 56-17-140 (small)
- Frame Material: Nylon
- Side Shields: 0.50mm Pb Lead Strips (optional)
- CE Certified: Yes
- Color Options: Silver, Black, Blue, Fuchsia, Brown (gold)
- Rx Options: Single Version, Bifocal, Progressive
- Min PD Value: 64 (regular) | 58 (small)
LENS ATTENUATION
| Lens Type | Lead Equivalence | Attenuation @ 100kVp | Attenuation @ 150kVp |
| 4B Semi Finished Lens - Thin |
0.75mm Pb |
98.3% |
96.4% |
| 4B Semi Finished Lens - Thick | 1.00mm Pb | 99.1% | 98.2% |
| 6B Semi Finished Lens - Thin | 1.05mm Pb | 99.2% | 98.6% |
| 6B Semi Finished Lens - Thick | 1.35mm Pb | 99.5% | 99.0% |
COMPLIANCE STANDARDS
Intended Use: Lead Glasses, Goggles, and fitovers ("Leaded Eyewear") are intended to be. worn by persons who are present in the examination of operating room during medical procedures requiring the use of x-ray/fluoroscopy. They are intended primarily to protect the eyes of the operator from x-ray radiation.
EN 61331-1:2014: (EU) Protective devices against diagnostic Medical X-ray radiation. (Determination of attenuation properties of material)
EN 61331-3:2014: (EU) Protective devices against diagnostic Medical X-ray radiation Part 3: Protective clothing, Eyewear and Protective Patient shields.
ASTM 2547-18: (USA) Standard test method for determining the attenuation properties in a. primary x-ray beam of materials used against radiation.
Protection Options: Lead (Pb) equivalent material (mm Pb) ≥ 0.50mm LE at 150kV
- PL75: Plano Lens 0.75mm (± 0.05) Pb equiv.
- PL50: Plano Lens 0.50mm (± 0.05) Pb equiv.
- PL75: Prescription Lens 0.75mm (± 0.05) Pb equiv.
- PL50: Prescription Lens 0.50mm (± 0.05) Pb equiv.
- SS: Lead strip side shield 0.50mm (± 0.05) Pb equiv.
- SS50: Lead strip side shield 0.50mm (± 0.05) Pb equiv.
- SS75: Lead strip side shield 0.75mm (± 0.05) Pb equiv.
MAINTAINING YOUR EYEWEAR
Storage: Leaded Eyewear should be stored in a protective case in normal room temperature and low humidity conditions when not in use.
Maintenance: Examine Leaded Eyewear frequently and before each use. Cease use and replace or repair immediately if the eyewear are chipped, pitted, or damaged. You may tighten screws/hinges if they loosen. Cleaning the Frame: use any standard eyewear cleaning cloth. Diluted soap and water usually will not discolor or damage the frame. Cleaning the Lenses: Never use ammonia or chlorine-based cleaners. Use of alcohol is permitted for lenses that do not posses an anti-reflective (A/R) coat. To clean A/R coat lenses, soak in diluted soap and water. DO NOT allow water to remain or dry onto the lenses as this can discolor/damage the lens. Wipe and dry using a Protech cleaning cloth or equivalent soft, lint free cloth.